Whitley H35 HEPA Hypoxystation
Product code : DWS.A07790
DWS.A07790The H35 HEPA Hypoxystation is a cell culture chamber ideal for those requiring HEPA filtration, as well as being able to accurately control oxygen, carbon dioxide, temperature and humidity<
You can contact us for more information via the button “quote request'
Whitley H35 HEPA Hypoxystation
Product description
The H35 HEPA Hypoxystation is a cell culture chamber ideal for those requiring HEPA filtration, as well as being able to accurately control oxygen, carbon dioxide, temperature and humidity
- Control O2 in 0.1% increments up to 20%; CO2 in 0.1% increments up to 15%; and relative humidity up to 80% - provides flexibility in your research and confidence in your results.
- Fitted with the unique Whitley Internal HEPA Filtration System. All the atmosphere passes through the filter every four seconds, which ensures the chamber environment is cleaned quickly. Placing the filters inside the workstation prevents the filter becoming saturated.
- Rather than use off-the-shelf filters, DWS has worked with one of the world's leading filter manufacturers to develop a filter specifically tailored to suit this Hypoxystation.
- Levels of atmospheric cleanliness inside the workstation exceeds the requirements of ISO 14644 Class 3. We can provide proof of atmospheric cleanliness.
- Fitted with a fan that circulates the internal atmosphere continuously - contributing to even temperature distribution and ensuring any internal particulate will be in motion.
- Has a removable front as standard to facilitate thorough cleaning and the transfer of bulk samples or larger pieces of equipment for use in the workstation.
- Built-in rapid airlock to ensure samples can be transferred into the workstation atmosphere as soon as possible (only takes 60 seconds).
- Colour, touch-screen control panel for ease of use and for visual display of parameters such as O2 level, CO2 level, temperature, humidity, and airlock cycle status.
- Fully integrated gas control avoids bulky add-on systems and reduces the need for additional bench space.
- CO2 Monitoring device fitted as standard to enable you to experiment using different concentrations of CO2.
An ideal solution for Seahorse Analyzer users is to purchase the H35 HEPA along with the new Whitley i2 Instrument Workstation. These two workstations can be connected via the Whitley Transfer Tunnel, enabling preparation of cell lines under hypoxic conditions and their transfer directly into the i2 without exposure to air.
Don Whitley Scientific provides the best possible service and maintenance support for this temperature controlled chamber. Please contact us to request a quotation.
Whitley H35 HEPA Hypoxystation
Product specifications
Brand
Don Whitley Scientific
Microscope
Available - please contact us for more information
Airlock Cycle Time (seconds)
60
Dimensions (w/d/h - mm)
1255 / 810 / 710
Remote Access
Fitted as standard (network access required)
Automatic Dehumidifier
Fitted as standard
HEPA
Fitted as standard
Internal Mains Socket
Available
Gas Supplies
CO2/Air/N2
Auto Sleeve Gassing
Available
Capacity (Litres)
300
Automatic Humidifier
Available
Extra Cable Glands
Available
Weight (lbs/kg)
264.5 / 120
Airlock Size
12 litres
Single Sample Entry
Available
Removable Front
Fitted as standard
Footswitch
Wireless
Touchscreen control
Yes
Data Logging
Available
CO2 Monitoring
Fitted as standard
Vacuum Take-off
Available
Porthole System
Manual, sleeved and oval ports
Storage Trays
Available
Whitley H35 HEPA Hypoxystation
Product documents
- Hypoxic Range Brochure
- HE04_ Whitley Internal HEPA Filtration - Bacteriological Testing
- HE08 - HEPA filtration of gas discharged from Whitley workstations_ bacteriological testing
- The Fastest Oxygen Control on the Market
- Whitley H35 HEPA Hypoxystation - Installation Document
- Whitley H35 HEPA Hypoxystation Fact Sheet
- Whitley HEPA Filtration System Literature
- Single Plate Entry Systems