Steam sterilizer HGD 113, steam generator, vacuum
Code produit : HMC.21007
HMC.21007High-end line, innovative top technology, with steam generator and vacuum pump coupled directly to the chamber
En cliquant sur le bouton "demande de devis", vous pouvez nous contacter pour plus d'informations.
Steam sterilizer HGD 113, steam generator, vacuum
Product Description
Innovative
Top technology
The latest generation - versatile with no compromises
Devices with steam generator and
vacuum pump coupled directly to the chamber
- Passive seal against vacuum and pressure
- Dynamic multi-directional inflow of steam
- Permanently available steam
- Segmented jacket heating
- No heating elements in the chamber
- Extremely short heating times
- Preheating program
- Pyrolytic self-cleaning program
- Powerful vacuum pump
- Extremely short batch times
All HMC steam sterilizers are produced in modern manufacturing facilities using state of the art technology and are TÜV certified. All devices meet the EC Pressure Equipment Directive and the current European regulations and bear the CE label.
Effective vacuum
Pre and drying vacuum
for the sterilization of
- Fabrics
- Hollow parts
- Tubes
- Bottles
- Pipettes
- Glass containers
- All open-pore materials
- Mixed fractions in waste bags
- Filters
Documentation
Simple via USB or printer (can be integrated)
Fast and safe!
The HGD devices are equipped with the most effective vacuum pump.
2 pre-vacuum cycles are sufficient to remove all air and let steam penetrate.
This saves time and guarantees the highest level of effectiveness.
The duration of the drying phase with fractionated vacuum is adjustable
and can therefore be matched individually to the load characteristics
Twice as safe
The locking system is coupled with a pressure and temperature monitoring system. Sensors monitor the lid state. Unintended opening under pressure not possible by design. The opening temperature (important for liquid sterilization) can be preset to between 60-97 °C (temperature dependend lid opening). No electronics and no moving parts in the lid, therefore reliable and trouble-free. Maximum service life!
Easy of use
The low loading height and the smooth surface make for easy loading and unloading. There are no obstructing locks or edges on which the baskets could catch.
Illuminated display
The clear, bright membrane display shows all important parameters at a glance and is self-explanatory. Simple menu navigation allows adaptations for customized program. 12 programs are available, divided into 4 groups with 3 sub-groups each.
The optimum for each load.
- Sterilization of solids: With fast release of steam
- Sterilization of liquids: With variable natural cooling or with rapid cooling for thermolabile materials
- Sterilization of liquids: with subsequent reheat cycle
- AGAR program: processes AGAR and keeps it ready for use (thermostatic heating)
- Timer-controlled program: load in the evening - sterile and ready for removal in the morning down to the minute
- Waste sterilization: Mixed load with a low liquid content
- Automatic standby operation: Saves energy
The timer function lets you start the program with a turn-on delay in order to have it end at a precisely defined time.
GREEN OPERATION
Save resources
Green operation
Top-quality materials are used to manufacture the high-tech chambers with a reduced mass. Result is a long service life and short heating and cooling times. Ideal energy efficiency.
Spent steam condenser
The innovative spent steam condensation system, highly efficient and matched to the energy and volume, prevents odor and steam pollution. This ensures a pleasant indoor climate in the environment of the device.
The HG series: Sophisticated functionality
The HG series devices are divided into working capacities of 50/80/113/133 litres and set standards with regard to comfort and safety. The user will be pleased by the extremely simple operation using the self-explanatory membrane keyboard with LED program display. The locking systems for both series consist of a motorized lid and an innovative locking system. The lid opens at the push of a button: first it only opens a safe gap in order to relieve the remaining pressure, then automatically and finally all the way. Unintended opening is impossible. The user is protected against escaping steam.Starting the program as well as opening and closing of the autoclave chamber are triggered via a single key. The lid is lowered gently and locked automatically. The program starts automatically.
The lock operating principles
HG/HGS/HGD 113/133
The double V-seal gasket seals equally well against vacuum and pressure. It is embedded flush in the edge of the chamber and therefore protected against damage during loading and unloading. The seals for both types can be replaced by the user themselves in just a few minutes.
Steam sterilizer HGD 113, steam generator, vacuum
Spécifications du produit
Packaging Unit
1 unit
Brand
HMC Europe
Category
HG-serie
Weight
184 kg
Dimensions (w x d x h)
686 x 870 x 930
Power
AC 400 V 50/60 Hz 3 phase
Chamber volume (l)
131,7
Steam generator
yes
Vacuum
yes
Chamber dimensions (Ø x H)
500 x 576
Heating power (kW)
6,8
Operating temperature
105° - 135°
Warming temperature
45° - 60°
Max. operating pressure (bar)
2,5 bar
Sterilization time
variable, 1 to 300 minutes
Automatic start
variable, 1 minute to 7 days
Locking system
automatic
Conformity
CE, TÜV Bayern e.V./TÜV Rheinland
Vessel material
SUS304 (high-grade stainless steel)
Documentation
Integrated printer/USB (optional)
Safety systems
Electro-mechanical safety lock, safety valve, low-water interrupt, overpressure protection switch, overheating protection switch, circuit breaker, fault indicator (visual and audible)