Omni BR-Cryo Cooling Unit
Code produit : OMN.19-8005
OMN.19-8005Cooling system to protect heat sensitive samples during homogenization…
En cliquant sur le bouton "demande de devis", vous pouvez nous contacter pour plus d'informations.
Omni BR-Cryo Cooling Unit
Product Description
Omni Bead Ruptor 24 Cryo Cooling Unit
Cooling system to protect heat sensitive samples during homogenization
The BR-Cryo Cooling Unit is compatible with the Bead Ruptor 24 and is designed to prevent the increase of sample temperature during the homogenization process. Friction generated through the bead milling process generates heat that is imparted to the sample. In order to maintain a constant processing temperature during homogenization, the BR-Cryo passes chilled air (-100°C to -50°C) into the sample processing chamber and across the sample tubes to dissipate heat.
Samples can be maintained at a temperature ranging from 4°C to ambient thus protecting heat sensitive samples such as RNA and proteins.
The BR-Cryo can be used to maintain a constant temperature when extracting functional enzymes or volatile small molecules.
Intra-tube temperature monitoring during bead mill homogenization. 2 mL Omni reinforced tubes containing five 2.8 mm ceramic beads (Cat # 19-628) were filled with 1 mL H2O and processed at 7 m/s on the Bead Ruptor 24 for increasing time periods. Homogenization was performed with and without the use of the BR-Cryo cooling unit using liquid nitrogen as the cryo source. Intra-tube temperatures were monitored at each time point. |
Bead Ruptor 24 Cryo Cooling Unit Overview
- Compatible with the Bead Ruptor 24
- Adjustable cooling rate
- Compatible with liquid nitrogen or dry ice with ethanol
- BR-Cryo maintain sample temperatures from 4°C to ambient
- Prevent protein degradation
- Ensure RNA stability
- Prevent enzyme activation
Laboratory Requirements
- Compressed dry medical grade air
- Compressed air flow rate must be at 120 PSI
- Liquid nitrogen or dry ice in ethanol (liquid nitrogen is preferred)
Omni BR-Cryo Cooling Unit
Spécifications du produit
Packaging Unit
1 pce
Brand
Omni
Category
BR-Cryo Cooling Unit
Weight
7.2 kg (16 lbs)
Dimensions (w x d x h)
29.2 (high) x 34.3 (deep at base) / 20.3 (deep at top) x 22.9 wide cm
Warranty
One year
Air Source Pressure
120 PSI
Air Source Requirements
Compressed medical dry air (water < 5ppm)
Cooling Source
Liquid Nitrogen or dry ice in ethanol
Cryo Chamber Volume
1 L
Dry Ice Consumption
1 L of 0.5 L ethanol + 0.5 L dry ice per 90 minutes of homogenization time
Liquid Nitrogen Consumption
1 L per 5 min of homogenization time
Temperature Range
5°C above ambient to +50°C
Timer
programmable / continuous